PolyU Institutional Repository >
Applied Physics >
AP Journal/Magazine Articles >
Please use this identifier to cite or link to this item:
http://hdl.handle.net/10397/616
|
| Title: | Factors affecting the performance of the bimorph-based dilatometer for field induced strain measurement of polymer films |
| Authors: | Guo, Shishang Chan, Helen L. W. Zhao, Xingzhong Choy, Chung-loong |
| Subjects: | Polymer films Strain sensors Extensometers Piezoelectric transducers |
| Issue Date: | Mar-2003 |
| Publisher: | American Institute of Physics |
| Citation: | Review of scientific instruments, Mar. 2003, v. 74, no. 3, p. 1285-1291 |
| Abstract: | This article gives a detailed discussion about the design and performance of a bimorph-based dilatometer using different sensor heads and platforms both experimentally and theoretically. A current preamplifier in high-pass filter mode was used to optimize the signal and a protective circuit was introduced. The test results demonstrated that the best signal occurred when the lengths of stainless steel pin and bimorph were both 10.0 mm and that the resonant frequency was inversely
proportional to the square of the sensor length. Some suggestions for reducing the noise and improving the sensitivity were also provided. |
| Rights: | © 2003 American Institute of Physics. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics. The following article appeared in S. Guo et al., Rev. Sci. Instrum. 74, 1285 (2003) and may be found at http://link.aip.org/link/?rsi/74/1285. |
| Type: | Journal/Magazine Article |
| URI: | http://hdl.handle.net/10397/616 |
| ISSN: | 00346748 |
| Appears in Collections: | AP Journal/Magazine Articles
|