Title: Comparison between reflection-mode photoplethysmography and arterial diameter change detected by ultrasound at the region of radial artery
Authors: Wang, Congzhi
Zheng, Yong-Ping
Subjects: Arterial diameter
Bood pressure
Potoplethysmography
Reflection-mode photoplethysmography
Ultrasound
Wrist volume clamp method
Issue Date: Aug-2010
Publisher: Lippincott Williams & Wilkins, Inc.
Citation: Blood pressure monitoring, Aug. 2010, v. 15, no. 4, p. 213-219.
Abstract: Objective: Continuous monitoring of arterial blood pressure (ABP) can provide vital information on the cardiovascular system in clinical practice and research. To achieve this, methods have been developed to determine ABP continuously and noninvasively, including the wrist volume clamp method, which uses reflection-mode photoplethysmography (PPG) to indicate the volume change of the radial artery. However, this kind of indication is reportedly not accurate enough, and an in-vivo test for its accuracy is still lacking. Our aim was to fix this gap.
Methods: As ultrasound can detect the arterial diameter accurately; we have developed a sensor comprised a PPG sensor and an ultrasound transducer, which can collect reflection-mode PPG and A-mode ultrasound signals, simultaneously. Tests on 15 volunteers were conducted under the similar conditions of using wrist volume clamp method for a comparison between the PPG signal and the radial arterial diameter.
Results: It was observed that there were obvious differences between the shapes of these two signals, mainly between their drifting trend parts, with the pulsatile components of the signals matching very well. The mean root mean square value of the differences of the trend parts (0.287±0.072) was found to be much larger than that of the pulsatile parts (0.107±0.028).
Conclusion: The results indicated that using PPG signal as a reference for ABP measurement at the wrist region might not be accurate enough, and ultrasound has the potential for replacing it in the wrist volume clamp method for more accurate ABP determination.
Description: DOI: 10.1097/MBP.0b013e328338aada
Rights: © 2010 Wolters Kluwer Health|Lippincott Williams & Wilkins. This is a non-final version of an article published in final form in Wang, C., & Zheng, Y. (2010). Comparison between reflection-mode photoplethysmography and arterial diameter change detected by ultrasound at the region of radial artery. Blood Pressure Monitoring, 15(4), 213-219.
Type: Journal/Magazine Article
URI: http://hdl.handle.net/10397/4368
ISSN: 1359-5237
Appears in Collections:HTI Journal/Magazine Articles

Files in This Item:

File Description SizeFormat
BPMJ-2009-50 (Pre-PublishedVersion).pdfPre-published version800.04 kBAdobe PDFView/Open
Locate publisher version via PolyU eLinks



Facebook Facebook del.icio.us del.icio.us LinkedIn LinkedIn


All items in the PolyU Institutional Repository are protected by copyright, with all rights reserved, unless otherwise indicated.
No item in the PolyU IR may be reproduced for commercial or resale purposes.

 

© Pao Yue-kong Library, The Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong
Powered by DSpace (Version 1.5.2)  © MIT and HP
Feedback | Privacy Policy Statement | Copyright & Restrictions - Feedback